The Brain Stimulator V3

Discontinued

Device Overview

User type: Consumer

Current type: Direct current

Current ramp-up rate/current ramp up time: N/A

Display: LED indicator

Current options (mA): 0.5, 1.0, 1.5, 2.0

Max voltage (V): 12

Timer settings (minutes): 30 minutes – not adjustable

Issues:

  • 12V supply is below the minimum compliance voltage of 20V for tDCS; insufficient for tDCS.

  • When the device is turned off, the device is still running in the background—causing the battery to drain.

  • LED does not display the true output, no real time current level.

  • No ramp-up feature, can be uncomfortable for some users.